KCH602 ventilator for emergency and transportation

Registered member contact customer service Click to Login
Retail price:
0.0
$
Market price:
0.0
$
Number:
KCH602
Quantity
-
+
Stock:
0
Login to download
catalogue
Temporarily out of stock
1
COMPLEMENTARY GOODS:
There is currently no content to display
Please add data record on website background.
Product description
parameter
KCH602 is a new generation of ventilator developed and manufactured by Zibo Kechuang medical instrument CO.Ltd.The system control
is performed using a microcomputer.A liquid-crystal display interface and touch screen buttons are used,so the respiratory parameters are easy to
set up and intuitively displayed, which ensures that the patients are provided with proper inspiratory to expiratory ratio, respiratory rate and tidal
volume. High quality imported products are used in main parts of the machine,thus greatly enhancing its reliability and stability.Appropriate
functions can be selected according to the patient's actual condition. Reasonable equipping with Lithium-ion battery significantly improves the
regional energy use limitation, thus providing a bete treatment space for the patients.
Stable technical properties:
Thecontinuousadjustmentsoftheparameterssuchasrespiratoryrate
(RR), inspiratory to expiratory ratio(1E), tidal volume(VT)and time of
synchronized ventilation ensure the applications in various clinicalsymptoms.
Stable technical properties:
Thecontinuousadjustmentsoftheparameterssuchasrespiratoryrate
(RR), inspiratory to expiratory ratio(1E), tidal volume(VT)and time of
synchronized ventilation ensure the applications in various clinicalsymptoms.
Reliable breathing patterns:
Breathing patterns such as synchronized ventilation,mandatory ventilation and manual ventilation are designed to provide more selections
for medical staffs and provide more advantageous service for patients'comfort and health.
Advanced alarm system:
Thealarmingparametersincludinguperlimitofpressure,lowerlimitofpowerandpoweroutageprovidestronginsuranceformedicalstff
and patients. The mute seting reflects humanization design of the ecquipment.
We could not find any corresponding parameters, please add them to the properties table

Product performance and technical parameters of KCH602 emergency and transfer ventilator
1.1 Working conditions
1.1.1 Environment: temperature -18℃~50℃, relative humidity 15%~95%, atmospheric pressure 70kPa~110kPa
1.1.2 Power supply: built-in lithium battery, 7.2V
1.1.3 Gas source: compressed oxygen (oxygen cylinder or medical gas pipeline system), working pressure 280kPa ~ 400kPa, flow range 0 ~ 40L/min.
1.1.4 Working mode: continuous operation
1.2 Breathing Mode: Forced, Synchronized (Auxiliary), Manual
1.3 Air circuit performance
1.3.1 The maximum limit pressure of the gas circuit: 6kPa.
1.3.2 Minimum limit pressure of gas circuit: trigger pressure (below atmospheric pressure).
1.3.3 The maximum working pressure is the ultimate pressure, and the minimum working pressure is 0kPa.
1.3.4 Negative pressure (subatmospheric pressure) cannot be obtained during the exhalation phase.
1.3.5 The circulating pressure range is between the lower pressure limit and the upper pressure limit.
1.3.6 The oxygen concentration range is 45% and 100%.
1.4 Setting parameters
1.4.1 Tidal volume: 100~1000ml, continuously adjustable.
1.4.2 Respiratory rate: 6~60 BPM, continuously adjustable.
1.4.3 Breathing ratio: 1:1, 1:2, 1:3, 1:4, 1:5, 2:1, 2:3, 3:5, continuously adjustable.
1.4.4 Pressure upper limit: 2~6kPa, continuously adjustable.
1.4.5 Lower limit of pressure: 0~2kPa, continuously adjustable.
1.4.6 Synchronization time: 0.2~2s, continuously adjustable.
1.4.7 Trigger pressure: -2kPa~0kPa, continuously adjustable.
1.4.8 Inspiratory oxygen concentration: 45% and 100% are optional.
1.5 Measured parameters
1.5.1 Tidal volume.
1.5.2 Respiratory rate.
1.5.3 Airway pressure (displayed by liquid crystal bar).
1.6 Alarm
1.6.1 Pressure upper limit alarm: 2~6kPa, when the pressure is higher than the set value, the sound and light alarm, and the alarm has no delay.
1.6.2 Pressure lower limit alarm: 0~2kPa, when the pressure is lower than the set value, sound and light alarm, the alarm delay is not more than 10s.
1.6.3 Lower limit of power: when the battery voltage is lower than 6.5V, the sound and light alarm, and there is no delay in the alarm.
1.6.4 Power failure alarm: alarm immediately after power failure, alarm duration ≥ 120s, sound alarm, no delay in alarm.
1.6.5 Pause of alarm sound: Pause time ≤120s.
1.7 Electrical performance
1.7.1 Working power supply: lithium battery, DC 7.2V.
1.7.2 Charging power requirements: AC 220V 50Hz.
1.7.3 Battery charger: meet the requirements of GB9706.1-2007.
1.7.4 Fuse: The specification is F1AL250V, and the size is φ5×20mm.
1.7.5 Host power: 28VA.
1.7.6 Electrical safety: Comply with the relevant requirements of GB9706.1-2007, YY0709-2009, YY0600.3-2007 for Class II type B application equipment.
1.8 Dimensions of the host: 280×120×180mm.
1.9 Net weight: 2.5kg.
1.10 Transportation and storage: The normal temperature range for storage and transportation of this product is -40°C to +80°C, the relative humidity range is 5% to 95%, there is no corrosive gas and a place with good ventilation. Put, pay attention to the directional signs.

Related documents

There is currently no content to display
Please add data record on website background.
There is currently no content to display
Please add data record on website background.
NEW ARRIVALS:
High-frequency electrosurgical unit
High-frequency electrosurgical unit
Market price: Available for inquiry
Negative pressure suction
Negative pressure suction
Market price: Available for inquiry

shopping cart

noGoods
No goods

What are you looking for:

Copyright © 2010-2021 Qisheng ( shanghai ) Medical equipment Co.,Ltd  ICP:2551450111